DENTAL TURBINE
One of the most important factors affecting the working life of dental tips is their care, i.e. cleaning and lubrication.
TURBINE END CARE
It is a complex mechanical tool with a rotor speed of up to 300,000 revolutions per minute. It is 5 thousand. ob. for a second. This species is particularly sensitive to poor care and careless handling. Therefore, the following rules must be observed.
It is important that drill bits have strictly regulated sizes. The use of drills with a longer length or a larger diameter of the working part disturbs the balance of the rotor, which in turn leads to rapid wear of the bearings. When using drills with a diameter other than the indicated dimensions, the rotor ring sleeve lock wears out quickly. As a result, the drill bit begins to “pop” when the tool starts. To control the drill diameter, there are special drill calibrators that are presented in various trade organizations.
Air pressure is equally important. Pressure exceeding the nominal value increases the rotational speed of the rotor group, but at the same time increases the load on the tip bearings, which significantly reduces its service life.
Do not press the sleeve button while the drill is rotating. Removal should only be done after the drill has come to a complete stop. Proper lubrication in accordance with the manufacturer’s recommendations is extremely important.
When lubricating the dental turbine with spray, it is necessary to use a special conical nozzle, the spray is fed into the working (drive) air hole. Models with a quick connector are lubricated either by the connector itself or by special nozzles that mimic the shape of the quick connector.
The head of the tip needs to be wrapped in a clean napkin. Apply the spray within 2-3 seconds. After you finish smearing, you need to check the stain on the napkin. If dirty dots or other contaminants are detected, repeat the lubrication procedure.
Once a day, it is necessary to lubricate the collet lock by applying a spray to the hole for 1-2 seconds. It is recommended to clean the cooling nozzle from time to time using a special wire. After lubrication, the dental tip should be cleaned of any remaining oil before use .
Important: These recommendations are general in nature. If the turbine operating instructions contain instructions different from those given above, the instructions contained in the instructions must be strictly followed.